Multiparameter Water Quality Analyzer

Product Introduction

Multiparameter Water Quality Analyzer designed for measuring pH, temperature, dissolved oxygen, conductivity, turbidity, and other important parameters. This integrated system combines multiple measurement capabilities into a single unit, ensuring enhanced efficiency and ease of use for all operators. It features direct connectivity with smart electrodes and multiparameter electrodes, allowing for seamless integration. The K100S provides real-time display of measurement values and electrode status, facilitating quick and accurate assessments. It is suitable for a variety of application scenarios, including surface water, municipal wastewater, industrial effluent, drinking water, aquaculture,e etc, making it a reliable choice for comprehensive water quality monitoring.

Categories

Conventional Five Parameters

 

Performance

Indicator Requirements

Dissolved oxygen

pH

Conductivity

Turbidity

Temperature

Measurement Principle

Fluorescence method

Glass electrode method

Electrode method

IR90°Light scattering method

PT1000

Measurement Range

0 ~ 20mg/L

0~14

0 ~ 200,000 μS/cm

0 ~ 4000 NTU

0~55 ℃

Repeatability Error

± 0. 3 mg/L

± 0.1 pH

±1%

±3%

±0.5℃

Zero Drift

< 0.1 mg/L

N/A

±1%

±3%

N/A

Range Drift

± 0. 3 mg/L

N/A

±1%

±5%

N/A

Drift

N/A

± 0.1 pH

N/A

N/A

N/A

Response Time (s)

45

15

15

N/A

N/A

Temperature Compensation Error

± 0. 3 mg/L

± 0.1 pH

±1%

N/A

N/A

Linearity Error

N/A

N/A

N/A

±5%

N/A

 

 

Optional Parameters

Performance

Indicator Requirements

ORP

COD

Ammonia nitrogen

Chlorophyll

Blue-green algae

Measurement Principle

Platinum electrode method

UV254nm Law

Turbidity compensation by 546nm transmission method

Ion selective electrode method

Fluorescence method 460nm/680nm

Fluorescence method 590nm/650nm

Measurement Range

-2000 ~ 2000 mV

0 ~ 500mg/L

0 ~ 500NTU

0 ~ 100mg/L

0-500.0 μg/L (chlorophyll a); 0 ~ 1000 ppb (Rhodamine WT)

0 ~ 2,000 Kcells/mL

Accuracy

± 1 mV

±10%

±10%

±5%

±5%

Resolution

0.1

0.01

0.01

0.01

0.01

Response Time (s)

15

NC

15

30

30

 

  1. The analyzer uses a touch screen with a large viewing angle.
  2. The analyzer supports accessing with a password to ensure security
  3. The analyzer adapts the standard MODBUS communication protocol
  4. The analyzer could measure at most 6 different parameters simultaneously
  5. The analyzer supports RTU transmission mode for connecting digital sensors
  6. The analyzer uses more than two sets of relays that can be programmed for either activate alarms or control purpose
  7. The analyzer features an electrically isolated MODBUS RTU interface
  8. The analyzer features a 0/4-20 mA analog output with electrical isolation
  9. The analyzer is capable of storing data for over 90 days.
  10. The analyzer features built-in temperature compensation to ensure accurate measurements.

Product Categories